A Validated and Simplified RP-HPLC Method for Estimation of Ebastine from Bulk Drugs
Subscribe/Renew Journal
A simple, specific, accurate and precise reverse phase liquid chromatographic method was developed and validated for the estimation of Ebastine from Bulk drugs. A RP HiQ Sil C8 HS column having 250x4.6 mm ID and 5 μm particle size in isocratic mode, with mobile phase containing Methanol and 10 mM Ammonium dihyrogen orthophosphate (80:20 % v/v) was used at the flow rate of 1.0 ml/min. The effluents were monitored at 255 nm. The retention time of Ebastine was 5.9 min. The linearity of the method was good (r > 0998), as also were intra-day and inter-day precision (RSD) < 2 %). The method is validated for accuracy, specificity, limit of quantification, limit of detection, robustness and stability. The results show that proposed method can be successfully applied for the quantitative determination of Ebastine.