Open Access Open Access  Restricted Access Subscription Access
Open Access Open Access Open Access  Restricted Access Restricted Access Subscription Access

RP-HPLC Determination of Amitriptyline Hydrochloride in Tablet Formulations and Urine


Affiliations
1 Prof. of Inorganic and Analytical Chemistry, Al-Azhar University, Cairo, Egypt
2 Forensic Chemist, Medico-Legal Organization, Ministry of Justice, Cairo, Egypt
     

   Subscribe/Renew Journal


A new, rapid and sensitive reverse phase HPLC method was developed and validated for the determination of amitriptyline hydrochloride in tablet formulations and urine. The mobile phase used acetonitrile and water, (50 % v/v) adjust pH to 5 using phosphoric acid. The separation was achieved on C18 reversed-phase column (250 mm x 4 mm i.d.). The flow rate was 0.6 ml/min and UV detection at 254nm. The retention time for amitriptyline hydrochloride was 7.3 min. The calibration curve was linear in the range 0.5-3 μg/mL. The mean recovery for amitriptyline hydrochloride is 100.025 %. The assay was precise within day and between days. The method provided excellent sensitivity, recovery, accuracy and reproducibility in therapeutic or toxic concentrations. Common excipients do not interfere.


Keywords

Amitriptyline Hydrochloride, Reverse Phase HPLC and C18 Reversed-Phase Column.
Subscription Login to verify subscription
User
Notifications
Font Size


Abstract Views: 225

PDF Views: 1




  • RP-HPLC Determination of Amitriptyline Hydrochloride in Tablet Formulations and Urine

Abstract Views: 225  |  PDF Views: 1

Authors

Rabie S. Farag
Prof. of Inorganic and Analytical Chemistry, Al-Azhar University, Cairo, Egypt
Ashraf M. Ahmed
Forensic Chemist, Medico-Legal Organization, Ministry of Justice, Cairo, Egypt

Abstract


A new, rapid and sensitive reverse phase HPLC method was developed and validated for the determination of amitriptyline hydrochloride in tablet formulations and urine. The mobile phase used acetonitrile and water, (50 % v/v) adjust pH to 5 using phosphoric acid. The separation was achieved on C18 reversed-phase column (250 mm x 4 mm i.d.). The flow rate was 0.6 ml/min and UV detection at 254nm. The retention time for amitriptyline hydrochloride was 7.3 min. The calibration curve was linear in the range 0.5-3 μg/mL. The mean recovery for amitriptyline hydrochloride is 100.025 %. The assay was precise within day and between days. The method provided excellent sensitivity, recovery, accuracy and reproducibility in therapeutic or toxic concentrations. Common excipients do not interfere.


Keywords


Amitriptyline Hydrochloride, Reverse Phase HPLC and C18 Reversed-Phase Column.