A Validated and Simplified RP-HPLC of Metoprolol Succinate from Bulk Drugs
Subscribe/Renew Journal
A simple, specific, accurate, and precise reverse phase liquid chromatographic method (RP-HPLC) was developed and validated for the estimation of Metoprolol succinate from bulk drugs. A RP Spherisorb C-18 (Waters) 10μm column having 250×4.6 mm ID in isocratic mode, with mobile phase containing acetonitrile: methanol: 10 mM aqueous phosphate buffer (20:20:60%v/v) was used. The flow rate was 1.0 ml/min and effluents were monitored at 254 nm. The retention time of Metoprolol succinate was 5.1 min. The linearity of the method was good (r > 0.998), as also were intra-day and inter-day precision (RSD <2%).
The method was validated for accuracy, specificity, limit of quantification, limit of detection, robustness and stability. The results showed that proposed method is successfully applied for the quantitative determination of Metoprolol succinate in bulk drugs.
Keywords
Abstract Views: 263
PDF Views: 0