Open Access Open Access  Restricted Access Subscription Access

To Evaluate the Marginal Fit of Metal Copings Fabricated by Conventional Casting Procedure and Direct Metal Laser Sintering Technology-An in Vitro Study


Affiliations
1 Dept. of Prosthodontics, Bhojia Dental College and Hospital, Himachal Pradesh, India
 

Purpose: The purpose of this in vitro study was to evaluate marginal fit of cobalt- chromium (Co-Cr) copings fabricated by direct metal laser sintering system (DMLS) and conventional lost-wax technique (LW).
Materials and method: Forty tooth preparations were carried out over extracted mandibular molars. They were divided into two groups A and B of 20 each. For group A Co-Cr copings were fabricated by direct metal laser sintering (DMLS) and for group B by lost wax technique (LW). Glass-ionomer cement (GIC) was used to tack the copings over their preparations. Marginal fit was then evaluated directly under the stereomicroscope.
Results: The mean marginal gap of group A was 27.9±2.4 μm and group B was 40.4 μm. Statistical analysis using t-test showed highly significant difference (P>.05) between the marginal mean of the DMLS (group A) compared to LW (group B).
Conclusion: The DMLS copings demonstrated superior marginal fit compared to that of conventional Co-Cr casted copings.

Keywords

Cobalt Chromium, Direct Metal Laser Sintering, Porcelain Fused Metal Crowns.
User
Notifications
Font Size

  • Petteno D, Schierano G, Bassi F, Bresciano ME, Carossa S. Comparison of marginal fit of 3 different metal-ceramic systems: an in vitro study. Int J Prosthodont 2000; 13:405-408.
  • Hunter AJ, Hunter AR. Gingival margins for crowns: a review and discussion. Part 2. Discrepancies and configurations. J Prosthet Dent 1990;64:636-42.
  • Kim KB, Kim WC, Kim HY, Kim JH. An evaluation of marginal fit of three-unit fixed dental prostheses fabricated by direct metal laser sintering system. Dent Mater 2013;29(7):e91-6.
  • Bindl A, Mormann WH. Marginal and internal fit of all-ceramic CAD/CAM crowncopings on chamfer preparations. J Oral Rehabil 2005;32:441-7.
  • Valderhaug J, Birkeland JM. Periodontal conditions in patients 5 years following insertion of fixed prostheses. Pocket depth and loss of attachment. J Oral Rehabil 1976;3: 237-43.
  • Valderhaug J, Heloe LA. Oral hygiene in a group of supervised patients with fixed prostheses. J Periodontol 1977;48:221-4.
  • Lang NP, Kiel RA, Anderhalden K. Clinical and microbiological effects of subgingival restorations with overhanging or clinically perfect margins. J Clin Periodontol 1983;10:563-78.
  • Ucar Y, Akova T, Akyil M, Brantley W. Internal fit evaluation of crowns prepared using a new dental crown fabrication technique: laser-sintered Co-Cr crowns. J Prosthet Dent 2009;102:253-259.
  • Faucher RR, Nicholls JI. Distortion related to margin design in porcelainfused-to-metal restorations. J Prosthet Dent 1980;43:149-55.
  • Buchanan WT, Svare CW, Turner KA. The effect of repeated firings and strength on marginal distortion in two ceramometal systems. J Prosthet Dent 1981;45:502-6.
  • Huang Z,Lu Zhang, Zhu J, Zhang X. Clinical marginal and internal fit of metal ceramic crowns fabricated with a selective laser melting technology. J Prosthet Dent 2015;113:623-27.
  • Anusavice KJ. Phillips' science of dental materials. 11th ed. Phialdelphia: W.B. Saunders; 2003. p. 565, 584, 585.
  • Vandenbroucke B, Kruth JP. Selective laser melting of biocompatible metals for rapid manufacturing of medical parts. Rapid Prototyping J 2007;13:196-203.
  • Wu L, Zhu H, Gai X, Wang Y. Evaluation of the mechanical properties and porcelain bond strength of cobalt-chromium dental alloy fabricated by selective laser melting. J Prosthet Dent 2014;111:51-5.
  • Yang X, Xiang N, Wei B. Effect of fluoride content on ion release from castand selective laser melting-processed Co-Cr-Mo alloys. J Prosthet Dent 2014;112:1212-6.
  • Zeng L, Xiang N, Wei B. A comparison of corrosion resistance of cobaltchromium-molybdenum metal ceramic alloy fabricated with selective laser melting and traditional processing. J Prosthet Dent 2014;112:1217-24.
  • Quante K, Ludwig K, Kern M. Marginal and internal fit of metal-ceramic crowns fabricated with a new laser melting technology. Dent Mater 2008;24:1311-1315.
  • Ortorp OA, Jonsson D, Mouhsen A, Von Steyern PV. The fit of cobalt-chromium three-unit fixed dental prostheses fabricated with four different techniques: A comparative in vitro study. Dent Mater 2011;27:356-363.
  • Oyague RC, Sanchez-Turrion A, Lopez-Lozano JF, Suarez Garcia MJ. Vertical discrepancy and microleakage of laser-sintered and vacuum-castimplant-supported structures luted with different cement types. J Dent 2012;40(2):123-30.
  • Kim KB, Kim WC, Kim HY, Kim JH. Evaluation of the marginal and internal gap of metal-ceramic crown fabricated with a selective laser sintering technology: two- and three-dimensional replica techniques. J Adv Prosthodont 2013; 5(2): 179-186.
  • Xu D,Xiang N,Wei B. The marginal fit of selective laser melting- fabricated metal crowns: an in vitro study. J Prosthet Dent 2014;112:1437-1440.
  • Shillingburg HT. Fundamentals of fixed prosthosthodontics. 4th ed. USA: Quintessence publishing; 2012.p.149-164.
  • Att W, Komine F, Gerds T, Strub JR. Marginal adaptation of three different zirconium dioxide three-unit fixed dental prostheses. J Prosthet Dent 2009;101:239-247.
  • Bhaskaran E, Azhagarasan NS, Miglani S, Gajapathi B.Comparative Evaluation of Marginal and Internal Gap of Co-Cr Copings Fabricated from Conventional Wax Pattern, 3D Printed Resin Pattern and DMLS Tech: An In Vitro Study. J Indian Prosthodont Soc 2013; 13(3): 189-195.
  • Akova T, Ucar Y, Tukay A, Balkaya MC, Brantley WA. Comparison of the bond strength of laser-sintered and cast base metal dental alloys to porcelain. Dent Mater 2008;24(10):1400-4.
  • Beschnidt SM, Strub JR. Evaluation of the marginal accuracy of different all-ceramic crown systems after simulation in the artificial mouth. J Oral Rehabil 1999;26(7):582-93.
  • Nawafleh NA, Mack F, Evans J, Mackay J, Hatamleh MM. Accuracy and reliability of methods to measure marginal adaptation of crowns and FDPs: a literature review. J Prosthodont 2013;22(5):419-28.
  • McLean JW, von Fraunhofer JA. The estimation of cement film thickness by an in vivo technique. Br Dent J 1971;131:107-11.
  • Li JM, Wang WQ, Ma JY. Comparison of the clinical effects of selective laser melting deposition basal crowns and cobalt chromium alloy base crowns. Shanghai Kou Qiang Yi Xue 2014;23:350-3.

Abstract Views: 319

PDF Views: 101




  • To Evaluate the Marginal Fit of Metal Copings Fabricated by Conventional Casting Procedure and Direct Metal Laser Sintering Technology-An in Vitro Study

Abstract Views: 319  |  PDF Views: 101

Authors

Monica Sharma
Dept. of Prosthodontics, Bhojia Dental College and Hospital, Himachal Pradesh, India
Ajay Bansal
Dept. of Prosthodontics, Bhojia Dental College and Hospital, Himachal Pradesh, India
Sunny Panthi
Dept. of Prosthodontics, Bhojia Dental College and Hospital, Himachal Pradesh, India
Shefali S. Malik
Dept. of Prosthodontics, Bhojia Dental College and Hospital, Himachal Pradesh, India
Atulya Sharma
Dept. of Prosthodontics, Bhojia Dental College and Hospital, Himachal Pradesh, India

Abstract


Purpose: The purpose of this in vitro study was to evaluate marginal fit of cobalt- chromium (Co-Cr) copings fabricated by direct metal laser sintering system (DMLS) and conventional lost-wax technique (LW).
Materials and method: Forty tooth preparations were carried out over extracted mandibular molars. They were divided into two groups A and B of 20 each. For group A Co-Cr copings were fabricated by direct metal laser sintering (DMLS) and for group B by lost wax technique (LW). Glass-ionomer cement (GIC) was used to tack the copings over their preparations. Marginal fit was then evaluated directly under the stereomicroscope.
Results: The mean marginal gap of group A was 27.9±2.4 μm and group B was 40.4 μm. Statistical analysis using t-test showed highly significant difference (P>.05) between the marginal mean of the DMLS (group A) compared to LW (group B).
Conclusion: The DMLS copings demonstrated superior marginal fit compared to that of conventional Co-Cr casted copings.

Keywords


Cobalt Chromium, Direct Metal Laser Sintering, Porcelain Fused Metal Crowns.

References