Charge Amplifiers Calibration in Vibration Metrology at NPLI


Affiliations
1 Acoustics, Ultrasonic, Shock and Vibrations Standards, Apex Level Standards and Industrial Metrology Division, National Physical Laboratory, CSIR, New Delhi, India
     

   Subscribe/Renew Journal


The present work discusses the methodology and procedure followed at NPLI for calibration of charge amplifiers. The charge amplifiers are calibrated conventionally using a 1 nF standard capacitor and sine generator for gain measurements. An automated calibration system using B&K Calibration software in conjunction with a calibrated 1 nF precision adaptor controlled by Pulse Analyzer is also used for precision calibration of charge amplifiers for gain and phase measurements from 0.5 Hz to 20 kHz. The uncertainty achieved in gain calibration is reported in the present work.
User
Subscription Login to verify subscription
Notifications
Font Size

Abstract Views: 200

PDF Views: 1




  • Charge Amplifiers Calibration in Vibration Metrology at NPLI

Abstract Views: 200  |  PDF Views: 1

Authors

Naveen Garg
Acoustics, Ultrasonic, Shock and Vibrations Standards, Apex Level Standards and Industrial Metrology Division, National Physical Laboratory, CSIR, New Delhi, India
Omkar Sharma
Acoustics, Ultrasonic, Shock and Vibrations Standards, Apex Level Standards and Industrial Metrology Division, National Physical Laboratory, CSIR, New Delhi, India

Abstract


The present work discusses the methodology and procedure followed at NPLI for calibration of charge amplifiers. The charge amplifiers are calibrated conventionally using a 1 nF standard capacitor and sine generator for gain measurements. An automated calibration system using B&K Calibration software in conjunction with a calibrated 1 nF precision adaptor controlled by Pulse Analyzer is also used for precision calibration of charge amplifiers for gain and phase measurements from 0.5 Hz to 20 kHz. The uncertainty achieved in gain calibration is reported in the present work.