Effect of Thermal Strain, Induced by Cryogenic Cooling, on a High Homogeneity Superconducting Magnet for MRI Applications


Affiliations
1 Inter-University Accelerator Centre, New Delhi, 110 067, India
2 Delhi Technological University, New Delhi, 110 042, India
 

The heart of the modern whole body MRI scanners is a superconducting magnet producing the required magnetic field with high homogeneity. The superconducting magnet operates at 4.2 K (-268.95 °C) and thus is kept dipped in liquid helium to maintain its superconducting state. This cool down of the multi-coil magnet from room temperature to the LHe temperature generates thermal strains in the magnet structure which deforms the magnet. These deformations are found to affect the final magnetic field homogeneity and introduce artifacts in the MR images. We report our results on studies on the stresses induced in the bobbin and the coils consequent upon cooling the magnet to 4.2 K. The maximum von Mises stress in the coils is calculated to be 31.3 MPa while in the bobbin it comes out to be 60 MPa. We find that the cool down causes a relative movement of the coils which in turn degrades the field homogeneity from 5.5 ppm to 278 ppm. The centre field is found to increase by 60 G is caused by a reduction in the overall cross-section of the coils. The change in homogeneity is also analysed in terms of Legendre polynomials where we found that the relative displacements of the coils introduce odd order terms to the polynomial expansion terms which were not present in the original design.

Keywords

Superconducting Magnets, Thermal Strain, Field Homogeneity, Cryogenic Cooling.
User
Notifications
Font Size

  • Lvovsky Y & Jarvis P, IEEE Trans Appl Supercond, 15 (2005) 1317
  • Thekkethil S R, Kar S, Kumar M, Soni V, Suman N K, Sharma R G, Rastogi V & Datta T S, IEEE Trans Appl Supercond, 28:4 (2018) 441905
  • Kar S, Suman N, Thekkethil S R, Soni V, Kumar R, Saini SK, Sharma R G, & Datta T S, Indian J Cryog 44 (2019) 193.
  • https://trc.nist.gov/cryogenics/materials/materialproperties.htm (24 Aug 2022)
  • Brice H, Theiler G & Gradt T. IEEE Trans Appl Supercond, 22 (2012) 4400204
  • Huang C, Wu Z, Huang R, Wang W & Li L. IOP Conf Ser Mater Sci Eng, 279 (2017) 012002.
  • Iwasa Y, Case Studies in Superconducting Magnets: Design and Operational Issues, 2nd ed. (Springer New York) 1994
  • Opera-3d User Guide 18R2 (Cobham Technical Services, UK) 2016.
  • Lvovsky Y, Stautner E W & Zhang T, Supercond Sci Technol, 26 (2013) 93001.
  • Blasche M & Fischer D, Magnet Homogeneity and Shimming, (Siemens Healthcare, Germany) 2015
  • Thekkethil S R, Kar S, Suman N, Kumar M, Soni V, Sharma R G, & Datta T S, Indian J Cryog, 43 (2018) 187

Abstract Views: 150

PDF Views: 82




  • Effect of Thermal Strain, Induced by Cryogenic Cooling, on a High Homogeneity Superconducting Magnet for MRI Applications

Abstract Views: 150  |  PDF Views: 82

Authors

Sankar Ram Thekkethil
Inter-University Accelerator Centre, New Delhi, 110 067, India
Vikas Rastogi
Delhi Technological University, New Delhi, 110 042, India
Soumen Kar
Inter-University Accelerator Centre, New Delhi, 110 067, India

Abstract


The heart of the modern whole body MRI scanners is a superconducting magnet producing the required magnetic field with high homogeneity. The superconducting magnet operates at 4.2 K (-268.95 °C) and thus is kept dipped in liquid helium to maintain its superconducting state. This cool down of the multi-coil magnet from room temperature to the LHe temperature generates thermal strains in the magnet structure which deforms the magnet. These deformations are found to affect the final magnetic field homogeneity and introduce artifacts in the MR images. We report our results on studies on the stresses induced in the bobbin and the coils consequent upon cooling the magnet to 4.2 K. The maximum von Mises stress in the coils is calculated to be 31.3 MPa while in the bobbin it comes out to be 60 MPa. We find that the cool down causes a relative movement of the coils which in turn degrades the field homogeneity from 5.5 ppm to 278 ppm. The centre field is found to increase by 60 G is caused by a reduction in the overall cross-section of the coils. The change in homogeneity is also analysed in terms of Legendre polynomials where we found that the relative displacements of the coils introduce odd order terms to the polynomial expansion terms which were not present in the original design.

Keywords


Superconducting Magnets, Thermal Strain, Field Homogeneity, Cryogenic Cooling.

References