Open Access
Subscription Access
Open Access
Subscription Access
Method Development and Validation of Valsartan and Its Impurities by High Performance Liquid Chromatography
Subscribe/Renew Journal
Our main objective is to develop an accurate and precise HPLC method for the determination of valsartan and its impurities. An inertsil ODS-3v; (150 × 4.6) mm; 5 μm column is used for the Separation of drugs. The flow rate was maintained at 1.0 mL/min and the wave length used for detection was 230 nm. The linearity was observed in the range of 0.025-50μg/ml of spiked impurities in valsartan, impurity B and impurity C with a correlation coefficient of 0.990, 0.996 and 0.997 respectively. Linearity, accuracy, precision and robustness parameters for the suggested method were estimated for validation. The developed method can be utilized in the analysis of valsartan tablets.
Keywords
HPLC, Method Validation, Related Impurities, Entacapone, LOQ, LOD.
Subscription
Login to verify subscription
User
Font Size
Information
Abstract Views: 323
PDF Views: 5