Open Access Open Access  Restricted Access Subscription Access
Open Access Open Access Open Access  Restricted Access Restricted Access Subscription Access

Base Degradation Study and Method Development of Rivaroxaban by RP-HPLC in Bulk


Affiliations
1 Shree Dhanvantary Pharmacy College, Department of Quality Assurance, Kim, Surat, India
     

   Subscribe/Renew Journal


A simple, precise and accurate HPLC method has been developed and validated for assay of Rivaroxaban. An isocratic separation was achieved using a phenomenex C18 (250×4.6 mm, 5μm), 100°A particle size columns with a flow rate of 1 ml/min and using a PDA detector to monitor the elute at 250 nm. The mobile phase consisted of Methanol: Acetronitrile (50:50, v/v). The method was validated for specificity, linearity, precision, accuracy and robustness. The method was linear over the concentration range of 20-100 μg/ml (r2=0.99995). The Drug was subjected to base degradation. The method was found to be robust and suitable for routine analysis of Rivaroxaban. Degradation products resulting from the stress studies did not interfere with the detection of Rivaroxaban and the method is thus stability-indicating.

Keywords

Rivaroxaban, Oral Anticoagulant, 250 nm, Stability Indicating Method, Base Degradation.
Subscription Login to verify subscription
User
Notifications
Font Size


Abstract Views: 213

PDF Views: 1




  • Base Degradation Study and Method Development of Rivaroxaban by RP-HPLC in Bulk

Abstract Views: 213  |  PDF Views: 1

Authors

A. Kasad Pinaz
Shree Dhanvantary Pharmacy College, Department of Quality Assurance, Kim, Surat, India
K. S. Muralikrishna
Shree Dhanvantary Pharmacy College, Department of Quality Assurance, Kim, Surat, India

Abstract


A simple, precise and accurate HPLC method has been developed and validated for assay of Rivaroxaban. An isocratic separation was achieved using a phenomenex C18 (250×4.6 mm, 5μm), 100°A particle size columns with a flow rate of 1 ml/min and using a PDA detector to monitor the elute at 250 nm. The mobile phase consisted of Methanol: Acetronitrile (50:50, v/v). The method was validated for specificity, linearity, precision, accuracy and robustness. The method was linear over the concentration range of 20-100 μg/ml (r2=0.99995). The Drug was subjected to base degradation. The method was found to be robust and suitable for routine analysis of Rivaroxaban. Degradation products resulting from the stress studies did not interfere with the detection of Rivaroxaban and the method is thus stability-indicating.

Keywords


Rivaroxaban, Oral Anticoagulant, 250 nm, Stability Indicating Method, Base Degradation.