Open Access Open Access  Restricted Access Subscription Access
Open Access Open Access Open Access  Restricted Access Restricted Access Subscription Access

Development and Validation of an HPLC Method for Analysis of Levetiracetam in Human Plasma


Affiliations
1 Department of Pharmaceutical Sciences, Andhra University, Visakhapatnam-530003, India
     

   Subscribe/Renew Journal


A simple Reverse Phase liquid chromatographic (HPLC) method for the determination of Levetiracetam in human plasma has been developed. Separation was achieved on a Prontosil C8 (150×4.6 mm, 5 μm) column with mobile phase composition of [pH 3.8 buffer:acetonitrile] (Solution A): HPLC grade water (Solution B) in the ratio 60:40 v/v with ultra violet detection at 210 nm. Lower limit of quantitation was 200 ng/ml. Maximum between-run precision was 5.91%. Mean extraction recovery was found to be 100.02 to 103.06%. Stability study showed that after three freeze-thaw cycles the loss of three quality control samples were less than 10%. Samples were stable at room temperature for 48h and at -20° for 2 months. Before injecting onto HPLC system, the processed samples were stable for at least 6 h. The method was used to perform bioequivalence study in human.
Subscription Login to verify subscription
User
Notifications
Font Size


Abstract Views: 165

PDF Views: 0




  • Development and Validation of an HPLC Method for Analysis of Levetiracetam in Human Plasma

Abstract Views: 165  |  PDF Views: 0

Authors

G. V. H. Raju
Department of Pharmaceutical Sciences, Andhra University, Visakhapatnam-530003, India
S. Ganapathy
Department of Pharmaceutical Sciences, Andhra University, Visakhapatnam-530003, India
D. G. Sankar
Department of Pharmaceutical Sciences, Andhra University, Visakhapatnam-530003, India
P. Y. Naidu
Department of Pharmaceutical Sciences, Andhra University, Visakhapatnam-530003, India

Abstract


A simple Reverse Phase liquid chromatographic (HPLC) method for the determination of Levetiracetam in human plasma has been developed. Separation was achieved on a Prontosil C8 (150×4.6 mm, 5 μm) column with mobile phase composition of [pH 3.8 buffer:acetonitrile] (Solution A): HPLC grade water (Solution B) in the ratio 60:40 v/v with ultra violet detection at 210 nm. Lower limit of quantitation was 200 ng/ml. Maximum between-run precision was 5.91%. Mean extraction recovery was found to be 100.02 to 103.06%. Stability study showed that after three freeze-thaw cycles the loss of three quality control samples were less than 10%. Samples were stable at room temperature for 48h and at -20° for 2 months. Before injecting onto HPLC system, the processed samples were stable for at least 6 h. The method was used to perform bioequivalence study in human.