Open Access
Subscription Access
Open Access
Subscription Access
A Selective High Performance Liquid Chromatographic Method for Estimation of Catechin in Ayurvedic Taila Preparations
Subscribe/Renew Journal
A simple, sensitive, rapid and selective high-performance liquid chromatographic (HPLC) method has been developed and validated for the analysis of catechin in marketed Ayurvedic Taila (oil) formulations containing Acacia catechu. Chromatography of methanolic: 0.1% formic acid (7:3, v/v) extracts of these formulations was performed on C18 (5 μm×25 cm×4.6 mm i.d) column using isocratic mobile phase consisting of methanol:acetonitrile:water (40:15:45, v/v/v) containing 1.0% acetic acid at a flow rate of 0.5 ml/min and SPD-10 AVP photodiode array (PDA) UV-Visible detector. The analytical marker, catechin, was quantified at 279 nm. The retention time of catechin was about 3.27 min. The linear regression analysis data for the calibration plot showed a good linear relationship with correlation coefficient of 0.9988 in the concentration range of 15 to 90 μg/ml for catechin with respect to peak area. The limit of detection and limit of quantitation values were found to be 0.5 μg/ml and 1.7 μg/ml respectively. Repeatability of the method was found to be 0.62 RSD. Recovery values from 99.73 to 100.22% indicate excellent accuracy of the method. The developed HPLC method is accurate, precise, and cost-effective, and it can be successfully applied for the determination of catechin in marketed ayurvedic oil formulations containing Acacia catechu.
Keywords
Acacia catechu, HPLC, Oil, Quantitation.
Subscription
Login to verify subscription
User
Font Size
Information
Abstract Views: 234
PDF Views: 0