Open Access Open Access  Restricted Access Subscription Access
Open Access Open Access Open Access  Restricted Access Restricted Access Subscription Access

Q-Absorbance Ratio Spectrophotometric Method for Simultaneous Determination of Atenolol and Ivabradine HCl in Synthetic Mixture


Affiliations
1 Department of Quality Assurance, Shree Dhanvantary Pharmacy College, Kim, Surat, Gujarat, India
2 Department of Pharmacognosy, Shree Dhanvantary Pharmacy College, Kim, Surat, Gujarat, India
     

   Subscribe/Renew Journal


It describes Simultaneous estimation of simple, accurate, precise, robust and economical Q-Absorbance ratio spectrophotometric method for Atenolol and Ivabradine HCl in synthetic mixture.

Objective: To Delivered information related to Ivabradine HCl and Atenolol combination's analytical method. Both drugs in combination have not any analytical method done by other or nor published anywhere. So we developed and validated it which is novel method for drugs.

Method: Absorbance ratio method for the ratio of absorbance at two selected wavelengths, one which is an isoabsorbtive point and other being the λ-max of one of the two component. Atenolol and Ivabradine HCl show an iso-absorbtive point at 286.40nm in methanol. The second wavelength used is 276nm which is λ-max of Atenolol in methanol. So it is economic in nature. The linearity was obtained in the concentration range of 20-100μg/ml for Atenolol and 2-10μg/ml for Ivabradine HCl. The concentration of the drugs was determined by using ratio of absorbance at iso-absorbtive point and at the λ-max of Atenolol.

Result: This method is linear for both drugs; in range 2-10μg/ml for Ivabradine HCl and 20-100μg/ml for Atenolol found at λmax of Atenolol 276nm (R2=0.9990) and at Iso-absorptive point 286.40nm (R2=0.9998). % recovery for Ivabradine HCl found 100.47% and Atenolol 100.32%. And all validation parameter (Robustness, Ruggedness, Interday, Intraday) show %RSD >2%. And Limit of detection for Ivabradine HCl and Atenolol at λ1(maximum wavelength) and λ2(Iso-absorptive point) was found 0.309 and 0.181 respectively. % Assay for Ivabradine HCl and Atenolol found to be 100.58% and 100.13% respectively.

Conclusion: The method was successfully applied to pharmaceutical synthetic mixture which is considered in approved patent which show no interference. The result of analysis has been validated statistically and by recovery studies. So this method accurate, precise, robust, rugged and economic in nature.


Keywords

Atenolol, Ivabradine HCl, Absorbance Ratio Method, Iso-Absorbtivepoint.
Subscription Login to verify subscription
User
Notifications
Font Size


Abstract Views: 278

PDF Views: 2




  • Q-Absorbance Ratio Spectrophotometric Method for Simultaneous Determination of Atenolol and Ivabradine HCl in Synthetic Mixture

Abstract Views: 278  |  PDF Views: 2

Authors

Pooja A. Patil
Department of Quality Assurance, Shree Dhanvantary Pharmacy College, Kim, Surat, Gujarat, India
Hasumati A. Raj
Department of Quality Assurance, Shree Dhanvantary Pharmacy College, Kim, Surat, Gujarat, India
Gautam B. Sonara
Department of Pharmacognosy, Shree Dhanvantary Pharmacy College, Kim, Surat, Gujarat, India

Abstract


It describes Simultaneous estimation of simple, accurate, precise, robust and economical Q-Absorbance ratio spectrophotometric method for Atenolol and Ivabradine HCl in synthetic mixture.

Objective: To Delivered information related to Ivabradine HCl and Atenolol combination's analytical method. Both drugs in combination have not any analytical method done by other or nor published anywhere. So we developed and validated it which is novel method for drugs.

Method: Absorbance ratio method for the ratio of absorbance at two selected wavelengths, one which is an isoabsorbtive point and other being the λ-max of one of the two component. Atenolol and Ivabradine HCl show an iso-absorbtive point at 286.40nm in methanol. The second wavelength used is 276nm which is λ-max of Atenolol in methanol. So it is economic in nature. The linearity was obtained in the concentration range of 20-100μg/ml for Atenolol and 2-10μg/ml for Ivabradine HCl. The concentration of the drugs was determined by using ratio of absorbance at iso-absorbtive point and at the λ-max of Atenolol.

Result: This method is linear for both drugs; in range 2-10μg/ml for Ivabradine HCl and 20-100μg/ml for Atenolol found at λmax of Atenolol 276nm (R2=0.9990) and at Iso-absorptive point 286.40nm (R2=0.9998). % recovery for Ivabradine HCl found 100.47% and Atenolol 100.32%. And all validation parameter (Robustness, Ruggedness, Interday, Intraday) show %RSD >2%. And Limit of detection for Ivabradine HCl and Atenolol at λ1(maximum wavelength) and λ2(Iso-absorptive point) was found 0.309 and 0.181 respectively. % Assay for Ivabradine HCl and Atenolol found to be 100.58% and 100.13% respectively.

Conclusion: The method was successfully applied to pharmaceutical synthetic mixture which is considered in approved patent which show no interference. The result of analysis has been validated statistically and by recovery studies. So this method accurate, precise, robust, rugged and economic in nature.


Keywords


Atenolol, Ivabradine HCl, Absorbance Ratio Method, Iso-Absorbtivepoint.