Open Access Open Access  Restricted Access Subscription Access
Open Access Open Access Open Access  Restricted Access Restricted Access Subscription Access

Development and Validation of a Simple and Rapid RP-HPLC Method for the Simultaneous Estimation of Amlodipine and Atorvastatin in Tablet Dosage Form


Affiliations
1 Department of Pharmaceutical Chemistry, P. E. S’s Rajaram and Tarabai Bandekar College of Pharmacy, Farmagudi, Goa - 403401, India
     

   Subscribe/Renew Journal


A simple and rapid reversed phase high performance liquid chromatographic method has been developed for the simultaneous estimation of Amlodipine and Atorvastatin in tablet dosage form. The separation was achieved on a Waters Symmetry C-18 column, (150mm × 3.9mm × 5μ particle size) using Acetonitrile: KH2PO4 buffer (pH 5.8) in the proportion of 45:55, v/v as mobile phase, at a flow rate of 1.0 mL/min. Detection was carried out at 254nm using VW detector. Amlodipine and Atorvastatin gave a retention time of 2.2±0.2 min and 3.2±0.2 min respectively. The developed method was validated for specificity, linearity, accuracy, precision and robustness as per ICH guidelines. The method was found to be specific as there was no interference seen in blank and placebo injections at the retention time of the drugs. The calibration curves were found to be linear with r2 = 0.999 for both the drugs. The method was found to be accurate as the mean percent recovery of Amlodipine and Atorvastatin was within the acceptance criteria. The method was found to be precise with %RSD less than 2. The method was found to be robust as small, deliberate changes in chromatographic conditions did not alter the results of analysis significantly. The method was applied in analysis of marketed formulation and the % assay was found to be within limits.

Keywords

Amlodipine, Atorvastatin, RP-HPLC, Validation, Simultaneous Estimation, Tablet Dosage Form.
Subscription Login to verify subscription
User
Notifications
Font Size


  • Tripathi KD. Essentials of Medical Pharmacology. Jaypee Brothers medical publishers Ltd., New Delhi. 2014.
  • Kumbhar S.T, Jadhav S.D, Bhatia N.M, Bhatia M.S. A Development and validation of derivative spectrophotometric method for estimation of Atorvastatin calcium and Amlodipine besylate in tablet dosage form. International Journal of Pharmacy and Pharmaceutical Sciences. 2011; 3(4):195-197.
  • Devi R., Ramakrishna S. New spectrophotometric method for the simultaneous Estimation of Amlodipine Besylate and Atorvastatin Calcium in tablet dosage forms. International Journal of Pharmacy and Pharmaceutical Sciences. 2010; 2(4):215-219.
  • Chaudhary BG, Patel AB. Simultaneous Spectrophotometric Estimation of Atorvastatin calcium and Amlodipine besylate in Tablet dosage form. International Journal of Chem Tech Research. 2010; 2(1):633-639.
  • Moussa BA, El-Zaher AA, Mahrouse MA, Ahmed MS. Simultaneous determination of Amlodipine besylate and Atorvastatin calcium in binary mixture by Spectrofluorimetry and HPLC coupled with Fluorescence detection. Analytical Chemistry Insights. 2013; 4(8):107-115.
  • Vidhyadhara S, Sasidhar RLC, Satyaprasanna P, Aruna K, Aruna KG. RP-HPLC method development for simultaneous estimation of Atorvastatin and Amlodipine besylate in bulk and pharmaceutical dosage forms. Plegia Research Library. Der Pharmacia Sinica. 2015; 6(10):1-7.
  • Nataraj KS, Alekhya K, SurendraBabu B, Gnananath K, Bindu C. Analytical method development and validation of simultaneous estimation of Amlodipine besylate and Atorvastatin calcium by RP-HPLC method. Journal of Chemical and Pharmaceutical Research. 2015; 7(4):35-41.
  • Hafez HM, Elshanawany AA, Abdelaziz LM, Mohram MS. Development of a Stability-Indicating HPLC Method for Simultaneous Determination of Amlodipine Besylate and Atorvastatin Calcium in Tablets. Austin J Anal Pharm Chem. 2014; 1(6): 01-011.
  • Ubale M. RP-HPLC Method for Simultaneous Determination of Amlodipine Besylate and Atorvastatin Calcium. Application to Commercially available drug products. Int. Arch. App. Sci. Technology. 2014; 5(1): 22-27.
  • Alnowaiser MA. Simultaneous determination of Amlodipine and Atorvastatin in Caduet® tablets using HPLC. American Journal of Applied Sciences. 2013; 10(8):849-852.
  • Turabi ZH, Khatatbeh OA. Stability indicating RP-HPLC method development and validation for the simultaneous determination of Atorvastatin and Amlodipine in pharmaceutical dosage forms. Journal of Pharmacy Research. 2013; 2(12):8-13.
  • Eranki RV, Inti G, Jayaraman V, Vidiyala SR, Ramulu J. New Stability Indicating Method for Quantification of Impurities in Amlodipine and Atorvastatin calcium Tablets by validated HPLC. International Journal of Pharmaceutical Research Scholars. 2013; 2(4): 458-471.
  • Silvia I, Klara K, Lucia MD. Simultaneous determination of Atorvastatin and Amlodipine in industrial tablet by Apparent Content Curve and HPLC method. Acta Medica Marisiensis. 2013; 59(1):44-48.
  • Dinda SC, Desireddy RB, Kumar PJ, Chand SA, Rao PS, Kumar BA. Development and validation of RP-HPLC method for quantitative analysis of Amlodipine besylate in Pure and Pharmaceutical Formulation. Research J. Pharm. And Tech. 2013; 6(2):204-207.
  • Momin MA, Islam MM, Rahman KH, Anisuzzaman SM. Development and Validation of assay method of Amlodipine Tablet by HPLC. International Journal of Pharmaceutical and Phytopharmacological Research. 2012; 2(2):109115.
  • Sharma AK, Dharamsi A. Development and validation of RP-HPLC and spectrophotometric method for the simultaneous estimation of Atorvastatin and Amlodipine in pharmaceutical dosage forms. International Journal of Pharmaceutical Sciences And Research. 2012; 3(4):1202-1207.
  • Ahmed M, Manohara YN, Ravi MC. RP-HPLC method development and validation for simultaneous estimation of Atorvastatin calcium and Amlodipine besylate. International Journal of chem. Tech Research. 2012; 4: 337-345.
  • Masud AA, Rahman MS, Hasan M, Ripon MKH, Khan AR, Islam MR, Ahasan MM, Uddin MR. Validated RP-HPLC method for simultaneous estimation of Amlodipine besylate and Atorvastatin calcium in combined tablet dosage form. Bangladesh Research Publication Journal. 2011; 6(1):52-56.
  • Acharjya SK, Mathrusri AM, Koya S. Liquid chromatographic method for simultaneous estimation of Atorvastatin calcium and Amlodipine besylate in pharmaceutical dosage forms. International Journal of Pharma and Bio Sciences. 2010; 1(4):161-170.
  • Shah DA, Bhatt KK, Mehta RS, Baldania SL, Gandhi TR. Stability Indicating RPHPLC estimation of Atorvastatin calcium and Amlodipine besylate in Pharmaceutical Formulations. Indian Journal of Pharmaceutical Sciences. 2008; 70(6): 754-760.
  • Zaheer Z, Farooqui MN, Mangle AA, Nikalje AG. Stability indicating high performance liquid chromatographic determination of Atorvastatin calcium in Pharmaceutical dosage form. African Journal of Pharmacy and Pharmacology. 2008; 2(10):204-210.
  • Q2B Validation of Analytical Procedures. http://www.ich.org/fileadmin/ Public_Web_Site/ICH_Products/Guidelines/Quality/Q2_R1/Step4/Q2_R1__ Guideline.pdf (15 October 2015).

Abstract Views: 213

PDF Views: 0




  • Development and Validation of a Simple and Rapid RP-HPLC Method for the Simultaneous Estimation of Amlodipine and Atorvastatin in Tablet Dosage Form

Abstract Views: 213  |  PDF Views: 0

Authors

Celina Nazareth
Department of Pharmaceutical Chemistry, P. E. S’s Rajaram and Tarabai Bandekar College of Pharmacy, Farmagudi, Goa - 403401, India
Shefali Naik
Department of Pharmaceutical Chemistry, P. E. S’s Rajaram and Tarabai Bandekar College of Pharmacy, Farmagudi, Goa - 403401, India
Anwesha Dourado
Department of Pharmaceutical Chemistry, P. E. S’s Rajaram and Tarabai Bandekar College of Pharmacy, Farmagudi, Goa - 403401, India

Abstract


A simple and rapid reversed phase high performance liquid chromatographic method has been developed for the simultaneous estimation of Amlodipine and Atorvastatin in tablet dosage form. The separation was achieved on a Waters Symmetry C-18 column, (150mm × 3.9mm × 5μ particle size) using Acetonitrile: KH2PO4 buffer (pH 5.8) in the proportion of 45:55, v/v as mobile phase, at a flow rate of 1.0 mL/min. Detection was carried out at 254nm using VW detector. Amlodipine and Atorvastatin gave a retention time of 2.2±0.2 min and 3.2±0.2 min respectively. The developed method was validated for specificity, linearity, accuracy, precision and robustness as per ICH guidelines. The method was found to be specific as there was no interference seen in blank and placebo injections at the retention time of the drugs. The calibration curves were found to be linear with r2 = 0.999 for both the drugs. The method was found to be accurate as the mean percent recovery of Amlodipine and Atorvastatin was within the acceptance criteria. The method was found to be precise with %RSD less than 2. The method was found to be robust as small, deliberate changes in chromatographic conditions did not alter the results of analysis significantly. The method was applied in analysis of marketed formulation and the % assay was found to be within limits.

Keywords


Amlodipine, Atorvastatin, RP-HPLC, Validation, Simultaneous Estimation, Tablet Dosage Form.

References