Open Access Open Access  Restricted Access Subscription Access
Open Access Open Access Open Access  Restricted Access Restricted Access Subscription Access

Development and Validation of Bioanalytical Method for the Estimation of Carisoprodol in Human Plasma using LC-MS/MS


Affiliations
1 College of Technology, Osmania University, Hyderabad, Telangana, India
2 School of Pharmacy, Aanurag Group of Institutions, Ghatkesar, Ranga Reddy, India
     

   Subscribe/Renew Journal


Simple, rapid and highly sensitive liquid chromatography- tandem mass spectrometric (LC-MS/MS) assay method was developed for the determination of Carisoprodol in human plasma. Carisoprodol methyl D3 was used as an internal standard (IS). The method employed 250μl of human plasma for sample processing by a simple Liquid Liquid Extraction (LLE) technique. The processed samples were chromatographer on a Phenyl column by using a mixture of 10mM Ammonium format - Acetonitrile (15:85,v/v) as the mobile phase at a flow rate of 1.2ml/min. The calibration curve obtained was linear over the concentration range of 25-3000ng/ml with r2 > 0.99. Method validation was performed as per the FDA guidelines and the results met the acceptance criteria. The Selected Ion Monitoring (SIM) mode was used for quantification of ion transitions at m/z 261.3/176.1 and 264.4/179.2 for the analyte and the IS respectively. A run time of 2.0 min was used which made it possible to analyze more than 400 plasma samples per day, thus increasing the productivity.

Keywords

Carisoprodol, LC-MS/MS, Bionalytical Method, Liquid-Liquid Extraction, Internal Standard.
Subscription Login to verify subscription
User
Notifications
Font Size


Abstract Views: 289

PDF Views: 2




  • Development and Validation of Bioanalytical Method for the Estimation of Carisoprodol in Human Plasma using LC-MS/MS

Abstract Views: 289  |  PDF Views: 2

Authors

P. Shashikala
College of Technology, Osmania University, Hyderabad, Telangana, India
D. Sireesha
School of Pharmacy, Aanurag Group of Institutions, Ghatkesar, Ranga Reddy, India
B. Vasudha
School of Pharmacy, Aanurag Group of Institutions, Ghatkesar, Ranga Reddy, India

Abstract


Simple, rapid and highly sensitive liquid chromatography- tandem mass spectrometric (LC-MS/MS) assay method was developed for the determination of Carisoprodol in human plasma. Carisoprodol methyl D3 was used as an internal standard (IS). The method employed 250μl of human plasma for sample processing by a simple Liquid Liquid Extraction (LLE) technique. The processed samples were chromatographer on a Phenyl column by using a mixture of 10mM Ammonium format - Acetonitrile (15:85,v/v) as the mobile phase at a flow rate of 1.2ml/min. The calibration curve obtained was linear over the concentration range of 25-3000ng/ml with r2 > 0.99. Method validation was performed as per the FDA guidelines and the results met the acceptance criteria. The Selected Ion Monitoring (SIM) mode was used for quantification of ion transitions at m/z 261.3/176.1 and 264.4/179.2 for the analyte and the IS respectively. A run time of 2.0 min was used which made it possible to analyze more than 400 plasma samples per day, thus increasing the productivity.

Keywords


Carisoprodol, LC-MS/MS, Bionalytical Method, Liquid-Liquid Extraction, Internal Standard.